摘要: 制备了3-巯丙基三甲氧基硅烷-铜(MPTMS-Cu)/MCM-41分子筛修饰碳糊电极,研究了该修饰电极对NO-2电化学还原的 电催化作用,分析了修饰物含量、溶液pH对电化学反应的影响。以该电极为检测电极,设计制作了芯片毛细管电泳-电化学 检测系统,探讨了NO-2的芯片毛细管电泳检测方法。以50 mmol/L醋酸钠(pH 5.8)为电泳缓冲液、分离电压为-1.6 kV时 ,检测可在40 s内完成。纯水中NO-2检测的线性浓度范围为10.0~5000.0 μmol/L,检出限为4.0 μmol/L。

关键词: NO-2, 电化学 , 芯片毛细管电泳, 修饰碳糊电极

Abstract: Carbon paste electrode modified with 3-mercaptopropyltrimethoxysilane copper (MPTMS-Cu) encapsulated in molecular sieve MCM-41 was prepared. The electrocatalytic behavior of the modified electrode towards the reduction of nitrite was studied in detail, including pH-dependence and composition-dependence studies. A microchip capillary electrophoresis-electrochemical detection system with the modified carbon paste as electrode was fabricated. The application of the system for the detection of nitrite is discussed. The detection was finished within 40 s under the following conditions: 50 mmol/L sodium acetate buffer at pH 5.8, -1.6 kV running voltage. The peak current was linear with the concentration of nitrite over 10.0 μmol/L-5.0 mmol/L and the detection limit was 4.0 μmol/L in pure water.