Chinese Journal of Chromatography ›› 2019, Vol. 37 ›› Issue (10): 1124-1128.DOI: 10.3724/SP.J.1123.2019.03001

• Articles • Previous Articles     Next Articles

Rapid determination of ammonia nitrogen in sewage by headspace gas chromatography

PENG Rong1, LIU Hao1, CHAI Xinsheng1, JIANG Ran2   

  1. 1. State Key Lab of Pulp & Paper Engineering, South China University of Technology, Guangzhou 510641, China;
    2. Pearl River Water Resources Institute, Pear River Water Resources Commission, Guangzhou 510611, China
  • Received:2019-03-01 Online:2019-10-08 Published:2020-12-11
  • Supported by:
    Guangzhou Science and Technology Plan, a Major Project of Industry-University-Research Collaborative Innovation (No. 201704020038);Guangxi Special Project for Innovation-Driven Development (No. AA18118013-9).

Abstract: This paper reports a novel method for the rapid determination of ammonia nitrogen in sewage by headspace gas chromatography (HS-GC) through a chemical reaction. During the experiment, formaldehyde, initially in excess, was reacted with NH4+ in a headspace bottle to produce a strong acid with equal molarity. Thereafter, NaHCO3 was reacted with the newly generated acid to produce CO2. In the final step, the released CO2 was determined using a gas chromatograph equipped with a thermal conductivity detector. Good linear correlation between the ammonia nitrogen content and the gas chromatographic signals was achieved. The observed correlation coefficient was 0.999, and the limit of quantification (LOQ) was 0.786 mg/L (measured based on the mass concentration of N). The relative standard deviation of reproducibility was less than 2%, and the standard recovery was between 95% and 105%. Compared with Nassar's reagent method, the relative deviation of this method is within ±5%. This method has high detection efficiency and is suitable for the rapid batch detection of ammonia nitrogen in sewage.

Key words: headspace gas chromatography (HS-GC), ammonia nitrogen, sewage, chemical reaction, CO2

CLC Number: