Chinese Journal of Chromatography ›› 2013, Vol. 31 ›› Issue (12): 1228-1231.DOI: 10.3724/SP.J.1123.2013.10007

• Technical Notes • Previous Articles     Next Articles

Determination of lutein in infant formula milk powder using ultra-high performance liquid chromatography

WANG Lina1,2, HUANG Junrong1, ZHANG Li2, FENG Feng2, LING Yun2, CHU Xiaogang2, LI Hongliang1   

  1. 1. College of Life Science and Engineering, Shaanxi University of Science and Technology, Xi'an 710021, China;
    2. Institute of Food Safety, Chinese Academy of Inspection and Quarantine, Beijing 100123, China
  • Received:2013-10-11 Revised:2013-11-12 Online:2013-12-28 Published:2013-12-05

Abstract:

An ultra-high performance liquid chromatography (U-HPLC) method for the determination of lutein in the infant formula milk powder was developed. The sample was extracted with acetone and defatted using freezing centrifugation method. The U-HPLC separation was achieved using a YMC Carotenoid C30 column (150 mm×4.6 mm, 3 μm) with the mixture of methanol/methyl tert-butyl ether (70:30, v/v) as the mobile phase under isocratic elution. The flow rate was 0.5 mL/min and the column oven temperature was 25 ℃. The injection volume was 5 μL. It was detected on a photodiode array detector at a wavelength of 445 nm. The results showed that the linear range was 20-500 μg/L (r=0.999 9), and the limit of quantification was 20 μg/L. The mean recoveries of lutein varied from 97.9% to 104.4% spiked at 50, 250 and 2000 μg/kg. The established method is simple, accurate and sensitive for the rapid determination of lutein in infant formula milk powder.

Key words: infant formula milk powder, lutein, ultra-high performance liquid chromatography (U-HPLC)