Chinese Journal of Chromatography ›› 2023, Vol. 41 ›› Issue (2): 152-159.DOI: 10.3724/SP.J.1123.2022.03033
• Articles • Previous Articles Next Articles
HUANG Yongpeng, TANG Hui, MENG Xiangyan, ZHONG Hui, SONG Yunyang, CHEN Bo(
), ZOU Zhiyun(
)
Received:2022-03-20
Online:2023-02-08
Published:2023-02-01
Contact:
CHEN Bo, ZOU Zhiyun
CLC Number:
HUANG Yongpeng, TANG Hui, MENG Xiangyan, ZHONG Hui, SONG Yunyang, CHEN Bo, ZOU Zhiyun. Rapid and simultaneous determination of two immunosuppressants in whole blood by high performance liquid chromatography[J]. Chinese Journal of Chromatography, 2023, 41(2): 152-159.
Add to citation manager EndNote|Ris|BibTeX
URL: https://www.chrom-china.com/EN/10.3724/SP.J.1123.2022.03033
| No. | Column style | Model | Specification | Maximum operating temperature/ ℃ | Status |
|---|---|---|---|---|---|
| 1 | traditional | ZORBAX Eclipse XDB C18 | 250 mm×4.6 mm,5.0 μm,8 nm | 60 | used |
| 2 | traditional | ZORBAX Eclipse Plus C18 | 250 mm×4.6 mm,5.0 μm,9.5 nm | 60 | used |
| 3 | traditional | ZORBAX Eclipse Plus C8 | 250 mm×4.6 mm,5.0 μm,9.5 nm | 60 | new |
| 4 | traditional | ZORBAX Eclipse PAH | 250 mm×4.6 mm,5.0 μm,9.5 nm | 60 | new |
| 5 | traditional | InfinityLab Poroshell 120 EC C8 | 150 mm×4.6 mm,2.7 μm,12 nm | 60 | new |
| 6 | biological | ZORBAX 300SB C8 | 250 mm×4.6 mm,5.0 μm,30 nm | 80 | new |
Table 1 Basic information of the employed chromatography columns
| No. | Column style | Model | Specification | Maximum operating temperature/ ℃ | Status |
|---|---|---|---|---|---|
| 1 | traditional | ZORBAX Eclipse XDB C18 | 250 mm×4.6 mm,5.0 μm,8 nm | 60 | used |
| 2 | traditional | ZORBAX Eclipse Plus C18 | 250 mm×4.6 mm,5.0 μm,9.5 nm | 60 | used |
| 3 | traditional | ZORBAX Eclipse Plus C8 | 250 mm×4.6 mm,5.0 μm,9.5 nm | 60 | new |
| 4 | traditional | ZORBAX Eclipse PAH | 250 mm×4.6 mm,5.0 μm,9.5 nm | 60 | new |
| 5 | traditional | InfinityLab Poroshell 120 EC C8 | 150 mm×4.6 mm,2.7 μm,12 nm | 60 | new |
| 6 | biological | ZORBAX 300SB C8 | 250 mm×4.6 mm,5.0 μm,30 nm | 80 | new |
Fig. 6 Relationships between the column temperature and chromatographic efficiency of cyclosporine A and sirolimus on different chromatography columns
| Immunosuppressant drug | Linear range/(ng/mL) | Linear equation | r | LOD/(ng/mL) | LOQ/(ng/mL) |
|---|---|---|---|---|---|
| Cyclosporine A | 30-2000 | Y=0.053X+0.045 | 0.9976 | 10 | 30 |
| Sirolimus | 2-200 | Y=0.098X+0.538 | 0.9989 | 1 | 2 |
Table 2 Linear ranges,linear equations,correlation coefficients(r),LODs,and LOQs of cyclosporine A and sirolimus
| Immunosuppressant drug | Linear range/(ng/mL) | Linear equation | r | LOD/(ng/mL) | LOQ/(ng/mL) |
|---|---|---|---|---|---|
| Cyclosporine A | 30-2000 | Y=0.053X+0.045 | 0.9976 | 10 | 30 |
| Sirolimus | 2-200 | Y=0.098X+0.538 | 0.9989 | 1 | 2 |
| Immunosuppressant drug | Added/ (ng/mL) | Found/ (ng/mL) | Recovery/ % | RSD/ % |
|---|---|---|---|---|
| Cyclosporine A | 30 | 25.06 | 83.5 | 9.0 |
| 60 | 51.26 | 85.4 | 6.0 | |
| 300 | 269.01 | 89.7 | 3.2 | |
| Sirolimus | 2 | 1.94 | 97.0 | 6.7 |
| 4 | 3.83 | 95.8 | 4.2 | |
| 20 | 19.56 | 97.8 | 3.4 |
Table 3 Spiked recoveries and RSDs of cyclosporine A and sirolimus in a blank whole blood sample(n=5)
| Immunosuppressant drug | Added/ (ng/mL) | Found/ (ng/mL) | Recovery/ % | RSD/ % |
|---|---|---|---|---|
| Cyclosporine A | 30 | 25.06 | 83.5 | 9.0 |
| 60 | 51.26 | 85.4 | 6.0 | |
| 300 | 269.01 | 89.7 | 3.2 | |
| Sirolimus | 2 | 1.94 | 97.0 | 6.7 |
| 4 | 3.83 | 95.8 | 4.2 | |
| 20 | 19.56 | 97.8 | 3.4 |
|
| Viewed | ||||||
|
Full text |
|
|||||
|
Abstract |
|
|||||